  • Model (7850) and company (Agilent) Location (General Room2,F3)

    Superuser (Xinghui Liu), contacts (13264791600/wechat ID: wxid_rfxhe72x6uo22)

  • Brief description

    Mainly used for high-precision and high-sensitivity metal element analysis, covering ultra-trace to constant concentration ranges. It is applicable to environmental monitoring, food testing, biological sample analysis, material analysis and other fields, and meets the analysis needs of ultra-trace elements.

  • Specifications

    -Injection system is well-equipped 4 high-precision gas mass flowmeters

    -Integrated quartz torch tube, easy to disassemble

    -The detector adopts a pulse simulation dual-mode electron multiplier, with an ion count range of ≥0.1cps to 1×1010 CPS per second.

    -LOD(3/1 Sign):Be(9) ≤ 0.5 ppt; In(115) ≤ 0.1 ppt; Bi(209) ≤ 0.1 ppt