CHN
Energy and Environment

Electrochemical workstation - VMP-3e

Source:Institute of LQM      Author:熊杰    Published Date:2026-03-12    Views:


  • Model (VMP-3e) and company (Bio-Logic), Location (Energy Room,F2)

   Superuser (Xinghui Liu), contacts (13264791600/wechat ID: wxid_rfxhe72x6uo22)

  • Brief description

    It is used in fields such as new energy materials, batteries, solid-state batteries, fuel cells, hydrogen production through water electrolysis, materials science, and electrochemical synthesis. It can conduct efficient and accurate tests on electrodes, achieve the rapid processing capability of multiple samples of electrode materials, facilitate the rapid screening of new material research and development, and accelerate the research process.

  • Specifications:

    - It can measure 32 samples simultaneously

    - Voltage highest resolution: 40 mu V, Working voltage: 0-10V

    - Voltage test precision: < + / - 0.1%

    - The current range: plus or minus 10 μA, ~ 100 mA

    - Current highest resolution: 0.2 nA, Current testing accuracy: < 0.1%;

    - The EIS electrochemical impedance function: all 32 channel support (sequence measurement 10 KHZ);

    - The user can through the WIFI network remote control equipment for testing;


Hubei University of Technology